Candela Corporation

Contact information

  • Address: 530 Boston Post Road Wayland, MA 01778, USA
  • Telephone: (800) 733-8550 or (508) 358-7637
  • Fax: (508) 358-5569
  • website: http://www.clzr.com
  • email:

Founded in 1970, Candela Corporation develops, manufactures, and distributes aesthetic lasers and other medical devices.

Available devices

GentleLASE Plus

Specifications

  • Laser Type: Flashlamp excited long-pulse alexandrite laser
  • Wavelength755nm
  • Pulse Characteristics
  • - Rep Rate1Hz
  • - Duration3 msec
  • - Spot Sizes8mm, 10mm, 12mm, 15mm, 18mm diameter3x10mm elliptical
  • - Fluence Range
  • 8mm: 20-100 J/cm2
  • 12mm: 6-40 J/cm2
  • 15mm: 6-30 J/cm2
  • 18mm: 6-20 J/cm2
  • 3x10mm: 30-100 J/cm2* (optional)
  • Beam deliveryLens-coupled optical fiber with handpiece
  • Pulse ControlFinger switch / Foot switch
  • Dimensions(HxWxD)43" x 20" x 32" (109cm x 51cm x 81cm)
  • Weight300 lbs.
  • Electrical requirements230V/50/60Hz, single phase, 13A
  • Integrated Dynamic Cooling Device
  • ComponentsIntegrated controls, cryogen container, and handpiece with distance guage
  • CryogenHFC 134a
  • Control SystemSpray Duration User adjustable range: 20-100ms
  • Delay Duration User adjustable range: 20-150ms
  •  

    integrated DCD. Spray cryogen

    18mm spot size

    Gentlelase Plus can also be used for treating vascular lesions, specifically leg veins from 0.5 to 2.0mm in diameter.

GentleLASEü LE

  • Specifications
  • Laser TypeFlashlamp excited long-pulse alexandrite laser
  • Wavelength755nm
  • Pulse Characteristics
  • - Rep Rate
  • 1Hz
  • - Duration3 msec
  • - Spot Size10mm
  • - Fluence Range10-40 J/cm2
  • Beam deliveryLens-coupled optical fiber with handpiece
  • Pulse ControlFingerswitch/Footswitch
  • Dimensions(HxWxD) " x 20" x 32" (109cm x 51cm x 81cm)
  • Weight300 lbs. (136 kg)
  • Electrical requirements230V/50/60Hz, single phase, 13A
  • Integrated Dynamic Cooling Device
  • ComponentsIntegrated controls, cryogen container, and handpiece with distance guage
  • CryogenHFC 134a
  • Control SystemSpray Duration - User adjustable range: 20-100ms
  • Delay Duration - User adjustable range: 3, 5, 10-150ms